Datascope Passport 2
$1,150
The Datascope Passport 2 is a vital signs monitor that is intended for intra-hospital use. These Datascope patient monitors are 6 trace monitors that feature an incredibly intelligent display interface that allows the user to evaluate patient status at a glance. Data is presented, concisely, and logically. Waveforms are aligned to digital data for immediate interpretation. Graph and list displays are easily configured to personal preference. The 4 Auto-Set Alarms program on the Datascope Passport 2 is quickly configured and extremely easy to use. Best of all, personalized custom settings retained on the PCMCIA cards assure a super-fast setup. Passport 2 can monitor the following parameters: ECG, Masimo SET SpO2, Nellcor Axismart or OxiMax SpO2, Non-invasive Blood Pressure, Respiration Rate, and Temperature. Optional parameters are 3 lead or 12 lead ST analysis with adjustable ISO and J points, Arrhythmia analysis, Invasive Blood Pressure, Gases, Microstream CO2, and 12 lead ECG interpretation.
The Datascope Passport monitor can be mounted on a rolling stand, wall mount bracket, gas machine arm, bed-rail, or operated as a tabletop instrument. It can be mounted to a Gas Module.
- 10.4” color TFT display
- 3 to 6 waveforms with automatic expansion
- Auto-adjustable large numerics
- 3 or 5 lead ECG with ESU noise rejection, lead-selectable respiration, NIBP, Masimo SET motion-tolerant SpO2, and temperature (400 or 700 series YSI probes)
- ECG and IBP analog output available for synchronization to balloon pump
- 6-minute OXY-CRG trend for neonatal and pediatric applications Built-in power supply for direct AC connection
Dimensions
- Height: 9.5" (24.1 cm)
- Width: 11.9" (39.2 cm)
- Depth: 7.4" (18.8 cm)
- Weight: 14.3 lbs. (6.5 kg) with 2 batteries.; 11.4 lbs. (5.2 kg) without batteries.
Battery Requirements
- 2 sealed lead acid batteries, 12 Vdc @ 2.3A load, 2.3 AH each battery.
Run time
- 2 hours +/- 6 minutes from two fully charged batteries at 25°C with ECG SpO2 and 1 NIBP every 15 minutes, no recording, no CO2.
- Recharge time: 16 hours maximum.
Power
- AC voltage: 100-240 VAC +/-10% 50/60 Hz +/-3 Hz Single Phase.
- Power consumption: 60-72 watts, 0.6-0.3 Amps.
Heart Rate Meter
- Range: 30-300 bpm adult/ Pediatric, 30-350 bpm Neonate.
- Accuracy: +/-3bpm or +/-3% at 30-250 bpm whichever is greater, +/-5% at 251-350 bpm.
IBP
- Range: Sys/Dia/Mean: 30- +300 mmHg.
- Accuracy: +/- 2 mmHg or 2% whichever is greater.
- Scale: -10 to 10, 0-20, 0-40, 0-160, 0-225, 0-320, 60-140 mmHg.
Temperature
- Scale: Selectable C° or F° Range: 15° to 45° C/ 59° to 113° F
- Accuracy: +/- 0.1°C (15° C to 45° C); +/- 0.2° F (59° to 113° F) exclusive of probe errors.
Respiration
- Range: 4 to 200 bpm.
- Accuracy: +/-2% or 2 breaths per minute whichever is greater from 4 to 150 bpm, +/- 4% from 151 to 200 bpm.
CO2 (Microstream)
- Range: 0-99 mmHg.
- Accuracy: 0-38 mmHg +/-2 mmHg, 39-99 mmHg +/-5% +0.08% for every 1 mmHg above 40 mmHg.
- Respiration rate: 0-150 bpm.
- Respiration accuracy: +/- 1 breath per minute from 0-40 bpm, +/-2 bpm from 41-70 bpm, +/-3 bpm from 71-100 bpm, +/-5 bpm from 101-150 bpm.
- ECG Leads: I, II, III, aVR, aVL, aVF, V.
Pulse Oximetry
- SpO2 saturation accuracy with no motion conditions.
- Adult / pediatric: 70 to 100% +/-2 digits SpO2, Neonate: 70 to 100% +/-3 digitsSpO2.
- SpO2 saturation accuracy during motion.
- Adult / pediatric / neonate: 70-100% +/-3 digits SpO2.
NIBP
- Technique: Oscillometric.
- Connector type: Rectus.
- Systolic range: adult: 55-235 mmHg, pediatric: 55-160 mmHg, neonate: 45-120 mmHg.
- Diastolic range: adult: 30-220 mmHg, pediatric: 30-150 mmHg, neonate: 20-100 mmHg.
- Systolic accuracy: mean error less that +/- 5 mmHg, standard deviation less than +/- 8 mmHg.
- Diastolic accuracy: mean error less that +/- 5 mmHg, standard deviation less than +/- 8 mmHg.
- Pulse rate range: adult/pediatric: 35 to 245 bpm, neonate: 70 to 245 bpm.
- Pulse rate accuracy: +/-3 bpm or 3% whichever is greater.